| Size (HxWxD): |
51 x 28 x 33 in
(130 x 72 x 84 cm) |
| Weight: |
(empty) 159 lb (72
kg) |
| Bottle
configurations: |
12 different
configurations available. 24 bottles, 1-liter PP or
350 ml glass; 4 bottles, 10-liter PE or glass; 2
bottles, 10-liter PE or glass; 1 bottle, 20-liter PE
or glass; 1 bottle, 10-liter PE or glass; 24 ProPaks,
1-liter wedge-shaped; 1 ProPak, 10-liter round |
| Refrigerator
body: |
Linear low-density
polyethylene (LLDPE) |
| Power
requirements: |
120 VAC, 60 Hz, or
240 VAC, 50 Hz (specify) |
| Operational
temperature: |
-20° to 120°F
(-29° to 49°C) |
| Pump |
| Intake suction
tubing: |
Length - 3 to 99
feet (1 to 30 m); Material - Vinyl or PTFE-lined vinyl;
Inside diameter - 3/8 inch (9 mm) |
| Pump tubing
life: |
Typically 1,000,000
pump counts. |
| Maximum suction
lift: |
28 feet (8.5 m) |
| Typical
repeatability: |
±5 ml or ±5% of
the average volume in a set. |
| Typical line
velocity @ Head height (with 3/8-in suction line): |
3.0 ft./s @ 3 ft;
2.9 ft./s @ 10 ft; 2.7 ft./s @ 15 ft. (0.91 m/s @ 0.9
m; 0.87 m/s @ 3.1 m; 0.83 m/s @ 4.6 m) |
| Liquid presence
detector: |
Non-wetted,
non-conductive sensor detects when liquid sample
reaches the pump to automatically compensate for
changes in head heights. |
| Controller |
| Enclosure
rating: |
NEMA 4X, 6 (IP67) |
| Program memory: |
Non-volatile ROM. |
| Flow meter
signal inputs: |
5 to 15 volt DC
pulse or 25 millisecond isolated contact closure for
Isco flowmeters. 4-20 mA input for non-Isco flowmeters. |
| Digital alarms: |
4 programmable
outputs; 5V, 100 mA |
| Number of
composite samples: |
Programmable from 1
to 999 samples. |
| Internal clock
accuracy: |
1 minute per month,
typical |
| Software |
| Sample frequency: |
1 minute to 99
hours 59 minutes, in 1-minute increments. 1 to 9,999
flow pulses. |
| Sampling modes: |
Uniform time, flow
proportioned, flow paced. (Flow modes are controlled
by external flow meter signal.) |
| Programmable
sample volumes: |
10 to 9,990 ml in 1
ml increments. |
| Sample retries: |
If no sample is
detected, up to 3 attempts; user selectable. |
| Rinse cycles: |
Automatic rinsing
of suction line up to 3 rinses for each sample
collection. |
| Controller
diagnostics: |
Tests for RAM, ROM,
pump, display, and electrical components. |